Thermal desorption (APK720R) |
Valve oven temperature | 150°C |
Transfer line temperature | 180°C |
Concentration |
1st desorption temperature | 300°C |
Focusing temperature | −20°C |
Focusing time | 10 min |
Desorption |
Temperature | 300°C |
Desorption time | 3 min |
GC-MS (Agilent7820A-5977E MSD) |
Inlet |
Temperature | 250°C |
Flow rate | 1 mL/min |
Oven temperature | 35°C (20 min)5°C/min to 50°C (10 min)5°C/min to 100°C (10 min)5°C/min to 130°C (10 min)5°C/min to 185°C (0 min)(Total 80 min) |
MS |
Aux-1 temperature | 300°C |
MS source | 230°C |
MS quad | 150°C |